1.12 TECHNICAL SPECIFICATIONS
Enclosure:
in terblend plastic (= ASA + polycarbonate):
- resistant to alcohol, glycols, most oils and greases, diluted acids
and water.
- non-resistant to aromatic hydrocarbons, esters, ethers, ketones,
concentrated mineral acids, ammonia gas and its dilutions.
Polyester front panel.
Aluminum varnish rear panel.
Flat keyboard with metal dome tactile response.
Dimensions:
Width 227 mm, depth 132 mm, height 77 mm without stand (87
mm with stand).
The stand is adjustable:
vertical or inclined at 13°.
A complementary base (supplied with the unit) allows an inclination
of 35°.
Degree of IP protection:
IP50 (according to IEC 529)
Weight of unit:
1.3 kg (2.9 lb)
Accuracy of measurement for D100S + probes:
interchangeability of probes and unit guaranteed as follows:
Probe type
D100S error
Probe error
Mean error
P5
1.0 µm
2.0 µm
2.2 µm
P10
1.2 µm
1.0 µm
1.6 µm
P25
1.5 µm
1.5 µm
2.1 µm
P50
3 µm
4 µm
5 µm
Repeatability (+/- 2s):
- for P5: 0.3 µm
- for P10: 0.2 µm
- for P25: 0.2 µm
- for P50: 0.4 µm
Operating temperature:
between + 5° and + 40°C
Storage temperature:
between -20° and + 45°C
Measuring frequency:
probe in
probe out
- for P10
between 170 and 205 measurements per sec.
- for P25/P50 between 130 and 205 measurements per sec.
The value measured is filtered digitally for the display, according to the resolution:
- 0.0001 mm or 0.00001 in:
approx. 3 readouts per sec. (= 3
analog outputs per sec.)
- 0.001 mm or 0.0001 in :
approx. 5 readouts per sec.
- 0.01 mm or 0.001 in :
approx. 12 readouts per sec.
- 0.1 mm or 0.01 in :
approx. 15 readouts per sec.
In Min/Max mode there is no filtering and thus maximum conversion speed is obtained.
Display:
LCD, STN type (Super Twisted Nematics) graphics 128 X 64
dots.Viewing area 66 X 33 mm. CCFL backlighting.
59