1.12 TECHNICAL SPECIFICATIONS

Enclosure:

in terblend plastic (= ASA + polycarbonate):
- resistant to alcohol, glycols, most oils and greases, diluted acids
and water.
- non-resistant to aromatic hydrocarbons, esters, ethers, ketones,
concentrated mineral acids, ammonia gas and its dilutions.

Polyester front panel.

Aluminum varnish rear panel.

Flat keyboard with metal dome tactile response.

Dimensions:

Width 227 mm, depth 132 mm, height 77 mm without stand (87
mm with stand).

The stand is adjustable:

vertical or inclined at 13°.
A complementary base (supplied with the unit) allows an inclination
of 35°.

Degree of IP protection:

IP50 (according to IEC 529)

Weight of unit:

1.3 kg (2.9 lb)

Accuracy of measurement for D100S + probes:

interchangeability of probes and unit guaranteed as follows:
Probe type

D100S error

Probe error

Mean error

P5

1.0 µm

2.0 µm

2.2 µm

P10

1.2 µm

1.0 µm

1.6 µm

P25

1.5 µm

1.5 µm

2.1 µm

P50

3 µm

4 µm

5 µm

Repeatability (+/- 2s):

- for P5: 0.3 µm
- for P10: 0.2 µm
- for P25: 0.2 µm
- for P50: 0.4 µm

Operating temperature:

between + 5° and + 40°C

Storage temperature:

between -20° and + 45°C

Measuring frequency:

probe in

probe out

- for P10

between 170 and 205 measurements per sec.

- for P25/P50 between 130 and 205 measurements per sec.

The value measured is filtered digitally for the display, according to the resolution:

- 0.0001 mm or 0.00001 in:

approx. 3 readouts per sec. (= 3
analog outputs per sec.)

- 0.001 mm or 0.0001 in :

approx. 5 readouts per sec.

- 0.01 mm or 0.001 in :

approx. 12 readouts per sec.

- 0.1 mm or 0.01 in :

approx. 15 readouts per sec.

In Min/Max mode there is no filtering and thus maximum conversion speed is obtained.

Display:

LCD, STN type (Super Twisted Nematics) graphics 128 X 64
dots.Viewing area 66 X 33 mm. CCFL backlighting.

59