Bowers Group Sylvac D100S Readouts User Manual
Page 27

26
1.6.11 SETUP KEY
Allows input of parameters for the following functions: Stat, Class, Tol, Clear or Enter, Min/Max, Set Up, Preset, Store,
Print, Scan and Channel. The input of parameters is explained with each corresponding function key.
Input of general parameters of unit:
1/ press
2/ then
Following will be displayed:
To exit this menu or the followings without changes, press
or
3.1/ External contact 1 function: menu 1 allows you to define the external contact function (in general the foot pedal
supplied with the unit). When the unit is ready for measurements, it is thus possible to inhibit the keyboard and use only
the foot pedal.
3.1.1/ Transmission of values through the RS 232C and Centronics outputs according to the parameters specified for
the Print function. Each time the foot pedal is pressed, the measured value is transmitted.
This choice may be combined with the following other functions:
2: display hold/store. The display is held as long as the foot pedal is pressed, when foot pedal is released, the mea-
sured value is sent. The functions 5 (change channels), 6 (D110 command), 7 (scanning) can be combined
as well.
3: Min/Max initialization. Alternately, pressing the foot pedal initializes Min and Max registers and pressing it again
sends the measured value. The functions 2 (hold/store), 5 (change channels) can be combined as well.
4: display preset. Alternately, pressing the pedal presets the display, pressing it again sends the measured value.
5: change channel. The same action on the foot pedal sends the measured value and then changes to the next
channel.
6: control the lifting of probes with D110 unit. The probes are lifted as long as foot pedal is pressed. When foot
pedal is released, the probes will go down and the measured value is sent. A delay can be introduced between
descent and transmission of value (value 1 to 36000 = 1 sec to 10 hours). Function 7 (start scanning) can be
combined as well. The first time the pedal is pressed the probes will be lifted, the second time they will go down,
scanning starts (after the delay, if any) and the value is sent.
7: scanning starting. Scanning is carried out first, followed by data transmission. A delay between the release of the
foot pedal and the scanning can be entered.
If a combination of the second or 3rd function is not required, simply press
.
1: Foot sw.1
2: Foot sw. 2
3: Language -Sprache-
Langue
4: Keyboard lock
5: Sound off
6: Set date/time
7: Configuration