LEVER TYPE DIAL TEST

INDICATOR INSTRUCTIONS

For more information visit

www.baty.co.uk

precision measuring instruments

Lever Type Dial Test Indicators
– BL Series

Please carefully review these instructions to ensure proper and accurate
use of your new Test Indicator.

Mounting of Mounting Shaft

Slide mounting shaft onto dovetail which is best suited for measuring and
tighten (fig. 1). Mount indicator rigidly and as close as possible.

Measuring

For accurate measurements the axis of the contact point must be perpendi-
cular to the measuring direction (fig. 2). If this is not possible, it is necessary
to multiply the reading on the dial with a correction factor, which depends
on angle ∂ (fig. 3). The correction factor is negligible for angles below 15º.

Angle ∂:

15º

30º

45º

60º

Correction factor:

0.96

0.87

0.70

0.50

Maintenance

No special maintenance is required. The bearings of the indicator must not
be oiled
.

Example

Angle ∂ :

30º (estimated)

Reading on dial:

.015"

Measurement value:

.015" x .87 = .013"

fig. 3

fig. 1

fig. 2