Bowers Group Sylvac D100S Readouts User Manual
Page 17

16
3/ confirm input number with
Following display will appear:
The channel selected by default in 3/ will appear. A number from 1 to 64 can
be entered.
4/ confirm input number with
Following display will appear
- press 1 for analog probes
- press 2 for OptoRS instruments used with simplex cable (no adapter required)
- press 3 for OptoRS instruments used with duplex cable(no adapter required)
The value of the RS232 instrument (indicator, caliper, Microcal...) will be dis-
played continuously on the D100S. The calculation functions or the unit are
available: TOL, CLASS, PRESET, STORE, STAT…
Using one or more D104 units, 16 RS232 channels can be connected to
D100S unit.
5/ The first display will then appear and a new allocation can be started, or exit with Clear.
Example: using one D102 with 2 P10 probes for internal bore measurements. The first probe measures 3 kinds of bore
with corresponding presets and classifications, and the second measures 4 types of bore with corresponding presets
and tolerances. An independent measurement is required on the last channel with a Microcal micrometer used with a
simplex cable.
- Channels 1,2 and 3 are thus allocated to probe No. 1:
1—>1
2—>1
3—>1
- Channels 4,5,6 and 7 are allocated to probe No. 2:
4—>2
5—>2
6—>2
7—>2
- Channel 8 is allocated to input 8 of the RS232 simplex:
8—>8RS
Input 2
Attributions :
= Input : 2