beautypg.com

Bowers Group Sylvac D100S Readouts User Manual

Page 30

background image

1.6.12 MIN/MAX KEY

Allows the choice of displaying minimum, maximum, difference Max - min, or mean (max + min)/2 instead of normal
measurement. Entering this mode will automatically set minimum and maximum registers to measuring position.
Min/Max measurements therefore start from this point.

If the channel is configured for an OptoRS duplex instrument, this mode is automatically linked by remote command to
the instrument.

In this mode, the preset will be given on the displayed value, i.e. on the maximum, the minimum, the difference or the
mean value.

Choice of the displayed function:

1/ press

2/ then

The following will be displayed:

Select required displayed value (min, max, etc.).

If the current channel number is smaller than the scanning number, the unit
asks if you wish to move to the next channel to enter the Min/Max choice.

- return to normal mode:

- return to choice above:

If min, max, delta and mean are to be stored or classified (TOL or CLASS), and this for the same probe, simply allocate
this probe to different channels, each one with a different function (min, max, etc.).

Example:

Channel 1 = probe 1 (standard) in the MIN mode, with its own TOL

Channel 2 = probe 1 (SET UP - CHANNEL) in the MAX mode, with its own TOL

When working in MIN/MAX mode, the

et

keys have a special function:

= Set Min/Max registers to the current measuring value. All new Min/Max values start from this point.

= Change the displayed value: Maximum --> Minimum --> Difference --> Mean --> Maximum...

CHANNEL 1

1 : MAX
2 : MIN
3 :

∆ = MAX-MIN

4 : MEAN= (MAX+MIN)/2

29