beautypg.com

Merit Medical EmboCath Plus Infusion Microcatheter IFU User Manual

Page 5

background image

Information on packaging:

Symbol

Designation

Manufacturer: Name & Address

Use by date: year-month

Batch code

Catalogue number

Do not resterilize

Do not use if package is damaged

Keep away from sunlight

Keep dry

Do not re-use

Caution -

Refer to Instructions For Use

Sterilized using ethylene oxide

EC mark logo -

Notified body identification : 0459

Maximum Guidewire

Maximum pressure

Radiopaque marker

Remove the Microcatheter from the hoop. Repeat
flushing of hoop if catheter is difficult to remove.
Inspect it prior to use to verify that it is undamaged

The hydrophilic infusion catheter must remain hydrated in
order to maintain lubricity.

INSTRUCTIONS FOR USE

The EmboCath Plus Infusion Microcatheter should be
utilized coaxially over a steerable guidewire. Follow
the guidewire manufacturer’s instructions for use.
The catheter may be used with a guide catheter/
sheath with a minimum internal diameter of 0.041
in. (1.04mm). A rotating hemostatic valve used in
conjunction with the guiding catheter will provide a
fluid-tight seal around the infusion catheter. Introduce
the infusion catheter and wire assembly through the
rotating hemostatic valve. Carefully tighten the valve
around the hydrophilic infusion catheter to prevent
backflow, allowing some movement through the valve
by the hydrophilic infusion catheter.

Overtightening of a hemostatic valve may cause
damage to the catheter.

Advance the guidewire and hydrophilic infusion
catheter to a selected vascular site by alternately
advancing the guidewire and then tracking the
hydrophilic infusion catheter over the guidewire.

Note: To facilitate catheter handling, the proximal portion
of the hydrophilic infusion catheter does not contain
the hydrophilic coating. Greater resistance may be en-
countered when this portion of the hydrophilic infusion
catheter is advanced into the rotating hemostatic valve.

To infuse, completely remove the guidewire from
the hydrophilic infusion catheter. Connect a syringe
with infusate to the hydrophilic infusion catheter hub
and infuse as required. Do not inject any fluid if air
bubbles are visible.

Note: Do not exceed Flow Rates/PSI as described in the
Flow Rate Table.

Aspirate infusion catheter upon removal and discard.

CAUTION:

PX Only: Federal (USA) law restricts this device to use by or

on the order of a licensed physician.

CONTRAINDICATIONS:

There are no known contraindications for the use of infusion
microcatheters.

STORAGE:

Store the EmboCath Plus Infusion Microcatheter in a cool,
dark, dry area.

COMPATIBILITY:

The EmboCath Plus Infusion Microcatheter should be
used with steerable guidewires that are up to 0.025
inches in diameter.

The EmboCath Plus Infusion Microcatheter is
compatible with microspheres of size range 700-900
μm and smaller.

Catheters exhibited no detectable signs of material
degradation during bench testing when exposed

between 60 and 75 minutes to a mixture of the
following agents: 5mg mytomicin, 25mg doxorubicin,
50mg cisplatin, 0.25g fluorouracil, 5mL ethiodol, 5cc
sterile contrast and 1cc sterile water.

BioSphere Medical assumes no liability with respect to
instruments reused, reprocessed or re-sterilized, and makes
no warranties, expressed or implied, including, but not limited
to, merchantability or fitness for intended use with respect to
such instrument.

All serious or life threatening adverse events or deaths
associated with use of EmboCath Plus should be reported
to the Competent Authority of the country where it occurred
and to the device manufacturer.

3