FIGURE 3-11. SAMPLE WAVEFORM
1. After observing the precautions of PAR. 3.3.9.1, press
@
from the power-up screen to enter
the Saved Waveforms screen (Figure 3-9), highlight a waveform designated
Empty
and
press
@
to create a new waveform.
2. The initial waveform settings are displayed (see Table 3-11 and Table 3-12 for parameter
details. Use
!
to modify the initial settings for the waveform. For each parameter, use
$
to save the setting.
Name
CAP CHARG
Mode
VOLTAGE
Protection
Positive
10
Negative
10
Count
0
Segment
Type
Neg Ramp
Frequency (Hz)
50
Amplitude (p-p)
4
Offset
8
When all settings are complete, press
$
to save for power-up. This step creates a repeat-
ing negative ramp starting at +10V decreasing to +6V since it is a 4V p-p signal riding on an
8V offset. The frequency of 50 Hz establishes the time duration of 20 ms for the segment.
3. From the Waveform Settings screen, highlight segment NEG RAMP 50.0HZ 4.00 R and
press
@
to insert a new segment. Use
!
to modify the following settings of the new seg-
ment. For each parameter, use
$
to save the setting.
Type
Level
Period (sec)
0.002
Offset
0
Initial/Repeat
Initial