A.17 *tst? — self test query, Table a-4. built-in test error codes, A.18 *wai — wait-to-continue command – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.2.01 to 2.37 User Manual
Page 146: A.17, A.18, H a.18, arra
A-6
BIT 4886 022108
A.17 *TST? — SELF TEST QUERY
*TST?
Syntax:
*TST?
Returned value: 0 or 1 (0 = pass test, 1 = fail test)
Description:
Power Supply test.This query causes the power supply to do a self test and provide the controller
with pass/fail results. A 0 is returned if the unit passes the test. If the unit fails, a number from 1
through 1023 is returned to indicate the cause of the error. The test executes each of the subtests
even when any one fails. If any test fails an error code bit is set which is returned to the user. The error
codes returned are listed in Table A-4
A.18 *WAI — WAIT-TO-CONTINUE COMMAND
*WAI
Syntax:
*WAI
Response:
Description: Causes the power supply to wait until all previously issued commands and queries are com-
plete before executing subsequent commands or queries. This command can be used to guaran-
tee sequential execution of commands and queries. When all pending operations are complete (all
previous commands have been executed, changes in output level have been completed), the WAI
command is completed and execution of subsequent commands can continue.
TABLE A-4. BUILT-IN TEST ERROR CODES
TEST ACTIVE
ERROR CODES
MEANING
*tst? DIAG:TST?
BIT
VALUE
Yes
Yes
0
Pass (No error)
Yes
Yes
0
1
ROM error
Yes
Yes
1
2
RAM error
Yes
Yes
2
4
FLASH error -- Perform calibration to correct
Yes
Yes
4
16
Analog Communication error
No
Yes
5
32
Loop Back Test error
No
Yes
6
64
Max Voltage Output error
No
Yes
7
128
Min Voltage Output error