5 applying saved settings to the output (recall), 9 waveform generation, 1 protecting the load when accessing waveforms – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.2.01 to 2.37 User Manual
Page 84: Caution, Before pressing, Applying saved settings to the output (recall) -24, Waveform generation -24, Protecting the load when accessing waveforms -24, R. 3.3.9, R. 3.3.8.5

3-24
BOP HIPWR 022108
NOTE: If a value to be saved is beyond the range of the unit, the value saved will be limited to
the maximum value. If the Mode is changed, the numerical values for the main and
protection channels must be revised accordingly.
• To modify the parameter, press
!
. Highlight the desired choice from a list, or use the
keypad to enter a numeric value, then press
$
to save, or
%
to abort.
• To exit without changing the parameter, press
%
.
5. Continue to modify parameters as described in step 4 above. When all parameters have
been modified, press
$
to save for power-up. Press
%
to exit without storing the
changes.
NOTE: If the message CHANGES NOT ALLOWED is displayed, the memory location has been
remotely locked (see PAR. A.12 for details).
6. After the changes have been saved, the unit returns to the Saved Settings screen, where
the saved settings can now be applied to the output by pressing
@
.
3.3.8.4
COPYING PREVIOUSLY SAVED SETTINGS TO A NEW LOCATION
1. Press
!
from the power-up screen to enter the Saved Setups screen (Figure 3-8).
2. Use the
U
and
Y
keys to highlight the location to be copied, then press
$
.
3. Highlight an
Empty
location and press
$
to save to the highlighted location, or press
%
to
abort the copy.
3.3.8.5
APPLYING SAVED SETTINGS TO THE OUTPUT (RECALL)
1. Press
!
from the power-up screen to enter the Save/Recall menu.
2. Use the
U
and
Y
keys to highlight the desired memory location. When the desired location
is highlighted, press
!
to view the detailed settings before execution, or press
@
to apply
the settings to the output. Refer to PAR. 3.3.8.3 to modify a setting before execution.
3.3.9
WAVEFORM GENERATION
3.3.9.1
PROTECTING THE LOAD WHEN ACCESSING WAVEFORMS
CAUTION: BEFORE PRESSING
@
TO ENTER THE SAVED WAVEFORM SCREEN FROM
THE POWER-UP SCREEN, PERFORM THE FOLLOWING STEPS TO AVOID
UNEXPECTED VOLTAGES OR CURRENTS FROM PREVIOUS SETTINGS
BEING BRIEFLY APPLIED TO THE LOAD WHEN THE WAVEFORM IS FIRST
EXECUTED.
1. Use the MODE key to set the operating mode of the unit (voltage or current) to correspond
with the mode of the waveform.
2. Set the main channel to zero, and the protection channel to maximum. This ensures that the
waveform will start from zero. Otherwise, between pressing STANDBY to turn the output on,
and pressing
$
to execute the waveform, the previously programmed settings of the power
supply will be applied to the load until the waveform starts executing.