B.13 OUTPut[:STATe]

QUERY

OUTP?

Syntax:

Short Form: OUTP[:STAT]?

Long Form: OUTPut[:STATe]?

Return Value: (0 or 1)

Description: Indicates whether power supply output is enabled or disabled. Returns 0 if output disabled,

returns 1 if output enabled. Related Commands: OUTP.

B.14 OUTPut:MODE COMMAND

OUTP:MODE

Syntax:

Short Form: OUTP:MODE {ACTIVE | RESISTIVE | BATTERY}
Long Form: OUTPut:MODE {ACTIVE | RESISTIVE | BATTERY}

Description: Determines how the unit functions when output is turned off depending on the type of load in

use: active, resistive or battery. See PAR. 3.3.6 for description of ACTIVE, RESISTIVE and BAT-
TERY modes.

B.15 OUTPut:MODE?

QUERY

OUTP:MODE?

Syntax:

Short Form: OUTP:MODE?

Long Form: OUTPut:MODE]?

Return Value: ACTIVE | RESISTIVE | BATTERY

Description: Indicates whether power supply is configured for ACTIVE, RESISTIVE or BATTERY type loads.

See PAR. 3.3.6 for description of ACTIVE, RESISTIVE and BATTERY modes. Related Commands:
OUTP:MODE.

B.16 [SOURce:]FUNCtion:MODE COMMAND

FUNC:MODE

Syntax:

Short Form: FUNC:MODE {VOLT | CURR} Long Form: [SOURce:]FUNCtion:MODE {VOLT | CURR}

Description: Establishes the operating mode of the power supply. VOLT = Constant Voltage mode (CV). CURR

= Constant Current mode (CC). FUNC:MODE VOLT commands power supply to Voltage mode,
FUNC:MODE CURR commands power supply to Current mode. Commanded mode establishes
parameters (voltage or current) monitored for error conditions. Actual mode depends upon load condi-
tions. When commanded to Voltage mode, if load conditions cause the power supply to try to exceed
the current limit, the unit will flag an error condition. When commanded to Current mode, if load condi-
tions cause the power supply to try to exceed the voltage limit, the unit will flag an error condition.

B.17 [SOURce:]FUNCtion:MODE? QUERY

FUNC:MODE?

Syntax:

Short Form: FUNC:MODE?

Long Form: [SOURce:]FUNCtion:MODE?

Return value: 0 (voltage mode) or 1 (current mode)

Description: Returns the operating mode of the power supply. 0 = Constant Voltage mode (CV). 1 = Constant

Current mode (CC).

B.18 [SOURce:]FUNCtion:MODE:TRIGger COMMAND

FUNC:MODE:TRIG

Syntax:

Short Form: FUNC:MODE:TRIG {VOLT | CURR}
Long Form: [SOURce:]FUNCtion:MODE:TRIGger {VOLT | CURR}

Description: Establishes the operating mode of the power supply when a TRIGger command is sent. VOLT =

Constant Voltage mode (CV). CURR = Constant Current mode (CC). FUNC:MODE:TRIG VOLT com-
mands power supply to Voltage mode when a TRIGger command is sent, FUNC:MODE:TRIG CURR
commands power supply to Current mode when a TRIGger command is sent.

B.19 [SOURce:]FUNCtion:MODE:TRIGger? QUERY

FUNC:MODE:TRIG?

Syntax:

Short Form: FUNC:MODE:TRIG?

Long Form: [SOURce:]FUNCtion:MODE:TRIGger?

Return value: 0 = voltage mode or 1 = current mode

Description: Returns the operating mode which the power supply will enter when a trigger command is

sent. VOLT = Constant Voltage mode (CV). CURR = Constant Current mode (CC).

B.20 [SOURce:]LIST:CLEar COMMAND

LIST:CLE

Syntax:

Short Form: LIST:CLE

Long Form: LIST:CLEar>

Description: Clears all list entries by setting all pointers to 0. Also sets LIST:DIR to UP, LIST:GEN to DSEQ,

Related Commands: All LIST commands (See example, Figure B-2.)