B.114 SYSTem:REMote? QUERY
SYST:REM?
Syntax:
Short Form: SYST:REM?
Long Form: SYSTem:REMote?
Return Value: {1 | 0}
Description: Identifies whether unit it in remote mode (1) or local mode (0) during serial (RS 232) communi-
cation. See PAR. 3.5.6.3 and Figure B-5.
FIGURE B-5. SETTING THE UNIT TO REMOTE MODE VIA SERIAL (RS 232) PORT
B.115 SYSTem:SECurity:IMMediate COMMAND
SYST:SEC:IMM
Syntax:
Short Form: SYST:SEC:IMM
Long Form: SYSTem:SECurity:IMMediate
Description: Initializes all NVRAM variable to factory defaults. Empties all memory locations. See PAR. 3.5.2.2
and Figure 3-16 for special programming considerations.
B.116 SYSTem:SET COMMAND
SYST:SET
Syntax:
Short Form: SYSTem:SET {CM0 | CM1 | DCL0 | DCL1 | LF0 | LF1 | RO0 | RO1 | OUT0 | OUT1}
Long Form: SYSTem:SET {CM0 | CM1 | DCL0 | DCL1 | LF0 | LF1 | RO0 | RO1 | OUT0 | OUT1}
Description: Establishes Device Clear, Line Feed, OUTP command and Reset functions. Sending
SYST:SEC:IMM sets LF1, DC0, and RO0 (as if CM0 was sent). Requires MEM:UPD (PAR. B.11) to
save for power up. See PAR. 3.5.2.2 and Figure 3-16 for special programming considerations.
DCL0
Device Clear functions per IEEE 488.2 Standard. (No effect on the device (power supply),
only clears internal status registers.)
DCL1
Device Clear functions identical to *RST (Output set to 0V, voltage mode and output set to
OFF except if RO1 (see below) is set.)
LF0
Line Feed not provided upon empty buffer condition
LF1
Line Feed provided if buffer is empty and a read is performed.
RO0
*RST and power up set output to OFF, 0V, 0A, voltage mode.
RO1
*RST sets output to ON, 0V, 0A, voltage mode. Subsequent VOLT and CURR commands
are immediately present at the output without sending OUTP ON (OUTP OFF and OUTP
ON function normally to turn the output off or on.)
OUT0
OUTP command functions per PAR. B.12
OUT1
OUTP command functions as described in PAR. B.12, however sending VOLT or CURR
command also turns the output on.
CM0
Establishes DCL0, LF0, RO0 conditions described above. (SCPI 1997 Standard compli-
ance)
CM1
Establishes DCL1, LF1, RO1 conditions described above. (Compatible with software ver-
sions 1.2 and earlier) (See example, Figure B-6.)
B.117 SYSTem:SET? QUERY
SYST:SET?
Syntax:
Short Form: SYST:SET?
Long Form: SYSTem:SET?
Return Value: DCL
Description: Identifies functions established by SYST:SET command. (See example, Figure B-6.)
*IDN?
Unit responds with KEPCO,BOP 50-20,E1234,1.66 (typical).
outp?
Unit responds with 0 indicating output is off
syst:rem?
Unit responds with 0 indicating unit is in local mode.
syst:rem 1
Puts unit in remote mode.
outp on
Enables output
outp?
Unit responds with 1 (output on).
syst:rem 0
Unit set to local mode.