beautypg.com

2 viewing saved settings, Table 3-8. save/recall menu, Viewing saved settings -23 – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.2.01 to 2.37 User Manual

Page 83: Save/recall menu -23, R 3.3.8.2. t, R. 3.3.8.3, 8 ar

background image

BOP HIPWR 022108

3-23

3.3.8.2

VIEWING SAVED SETTINGS

1. Pressing

!

from the power-up screen displays the Saved Setups screen (Figure 3-8) con-

sisting of a list of locations 1 through 20 arranged in two columns of 10 rows. Use

T

and

R

keys to display additional columns. NOTE: Use the number keys as a shortcut, e.g., press-
ing

4

immediately highlights location 41, showing the column with locations 41 through 50.

2. Use the

U

and

Y

keys to highlight the desired memory location; When the desired location

is highlighted, press

!

. The details of the parameters listed in Table 3-8 are displayed.

3.3.8.3

SAVING SETTINGS AND ERASING OR MODIFYING PREVIOUSLY SAVED SETTINGS

1. Press

!

from the power-up screen to enter the Saved Setups screen (Figure 3-8).

2. Use the

U

and

Y

keys to highlight the desired memory location; To erase a highlighted

memory location, press

#

. The list will show erased locations as

Empty

.

3. Press

!

to see the details stored in the highlighted location.

4. Once the details are displayed, use the

U

and

Y

keys to highlight the desired parameter.

TABLE 3-8. SAVE/RECALL MENU

PARAMETER

CHOICES

(BOLD = Factory Default)

FUNCTION

NOTE: DEFAULT values for empty cells are the settings of the unit at the time the save/recall menu is entered.

MODE

VOLT

CURRENT

EXTERNAL

VOLT - Selects voltage mode. Voltage value determined by SETTING.
CURRENT - Selects current mode. Current value determined by SETTING.
EXTERNAL - Used for customized configuration where EXT VM/CM signal at pin
2 of I/O port controls mode; consult factory for details.

MAIN CHANNEL

REFERENCE

INTERNAL

EXTERNAL

EXTL REF LVL

INTERNAL: Any external reference applied to the Analog I/O port pin 11 is ignored
and the internal reference established by SETTING field.
EXTERNAL: The internal reference is ignored, and the external reference applied
to the Analog I/O port pin 11 is used (see PAR. 3.4.3.1).
EXTL REF LVL: Allows the gain of the BOP to be established by the maximum
output voltage or current introduced by the SETTING field. (see PAR. 3.4.3.2).

SETTING

nn.nnn or

nnn.nnn

(model and parameter

dependent)

Sets the active channel setting. E.g., if VOLT mode is selected, nnn.nnn is voltage
setting, if CURRENT mode is selected, nn.nnn is current setting. If EXTERNAL
mode is selected, SETTING is not available.

PROTECTION

MODE

INTERNAL

EXTERNAL

LESSER LIMIT

INTERNAL - Allows limits to be controlled by POSITIVE and NEGATIVE values.
EXTERNAL: Allows limits to be controlled by analog signals applied to the analog
port (see PAR. 3.4.4).
LESSER LIMIT: Allows protect limit to be automatically selected from either 1) the
external analog voltage applied to the Analog I/O port or 2) the value set in the
POSITIVE and NEGATIVE fields. Whichever limit has a lower absolute value
(closest to zero) has effect (see PAR. 3.4.4).

POSITIVE nn.nnn

or

nnn.nnn

This is the positive value for Voltage protect if CURRENT mode is selected, or the
positive value for Current protect if VOLTAGE mode is selected. If EXTERNAL
mode is selected, this field is not accessible.

NEGATIVE

nn.nnn or

nnn.nnn

This is the negative value for Voltage protect if CURRENT mode is selected, or the
negative value for Current protect if VOLTAGE mode is selected. If EXTERNAL
mode is selected, this field is not accessible.

OUTPUT

OFF

ON

OFF - Output disabled. The behavior of the unit when output is disabled depends
on the LOAD TYPE setting (see PAR.3.3.6).
ON - Output enabled.