B.85, Sour, B.86 – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.2.01 to 2.37 User Manual
Page 165: B.87, B.88, B.89, B.90, B.85, b.86, B.89, b.87, B.89, b.90

BOP-1K 022108
B-19
B.85 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:BOTH COMMAND
VOLT:PROT:LIM[:BOTH]
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:PROT:LIMit[:BOTH]
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIM[:BOTH
where
voltage.
Description: Establishes the maximum value possible for both the positive and negative protection limits for volt-
age. Requires MEM:UPD (PAR. B.11) to save for power up. See PAR. 3.5.2.2 and Figure 3-16 for spe-
cial programming considerations.
B.86 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit[:BOTH]? QUERY
VOLT:PROT:LIM[:BOTH]?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM[:BOTH]?
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit[:BOTH]?
Returns
Description: Identifies the maximum value possible for protection limits for voltage (positive, negative); maximum is
1% greater than rated output voltage.
B.87 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:NEGative COMMAND
VOLT:PROT:LIM:NEG
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:NEG
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:NEGative
where
voltage.
Description:Establishes the minimum (maximum absolute) value possible for the negative protection
limit for voltage. This may be set up to 1% greater than rated voltage. Requires MEM:UPD (PAR. B.11)
to save for power up. See PAR. 3.5.2.2 and Figure 3-16 for special programming considerations.
B.88 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:NEGative? QUERY
VOLT:PROT:LIM:NEG?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:NEG?
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:NEGative?
Returns:
Description:Identifies the minimum (maximum absolute) value possible for the negative protection limit
for voltage.
B.89 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive COMMAND
VOLT:PROT:LIM:POS
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:POS
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive
where
voltage.
Description:Establishes the maximum value possible for the positive protection limit for voltage. This
may be set up to 1% greater than rated voltage. Requires MEM:UPD (PAR. B.11) to save for power
up. See PAR. 3.5.2.2 and Figure 3-16 for special programming considerations.
B.90 [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive? QUERY
VOLT:PROT:LIM:POS?
Syntax:
Short Form: [SOUR:]VOLT[:LEV]:PROT:LIM:POS?
Long Form: [SOURce:]VOLTage[:LEVel]:PROTect:LIMit:POSitive?
Returns:
Description:Identifies the maximum value possible for the positive protection limit for voltage.