beautypg.com

GW Instek GDS-1000A Series User Manual User Manual

Page 95

background image

SAVE/RECALL

95

Detail

Contains the waveform amplitude and
time of each point (4k/1M/2M)
relative to the trigger point.

Fast

Only contains the waveform
amplitude data for each point
(4k/1M/2M).

Waveform type

CH1, 2

Input channel signal

Math

Math operation result (page59)

Storage location Internal

memory

The oscilloscope’s internal memory,
which can hold 15 waveforms.

External

SD/SDHC
card

An SD/SDHC card (FAT or FAT32
format) can hold practically an
unlimited number of waveforms.

Ref

A,

B

Two reference waveforms are used as
a buffer to recall a waveform in the
display. You have to save a waveform
into internal memory or an SD card,
then copy the waveform into the
reference waveform slot (A or B), and
then recall the reference waveform
into the display.

Waveform
Memory Depth

The memory depth is limited to 1 M points when
both channels are activated or 2M points when
only a single channel is activated. The signal must
be triggered /stopped to have access to the full
memory depth. Therefore when a signal is saved
the waveform will be automatically stopped if it is
not manually triggered /stopped first.
There are a number of conditions when all of the
available memory is not utilized due to a limited
number of different sample rates. This can be
caused by an un-triggered signal, or a time/div
setting that is too fast to display all the points on
screen.