GW Instek GDS-1000A Series User Manual User Manual
Page 95

SAVE/RECALL
95
Detail
Contains the waveform amplitude and 
time of each point (4k/1M/2M) 
relative to the trigger point. 
Fast
Only contains the waveform 
amplitude data for each point 
(4k/1M/2M). 
Waveform type
CH1, 2
Input channel signal
Math
Math operation result (page59)
Storage location Internal
memory
The oscilloscope’s internal memory, 
which can hold 15 waveforms. 
External
SD/SDHC 
card 
An SD/SDHC card (FAT or FAT32 
format) can hold practically an 
unlimited number of waveforms. 
Ref
A,
B
Two reference waveforms are used as 
a buffer to recall a waveform in the 
display. You have to save a waveform 
into internal memory or an SD card, 
then copy the waveform into the 
reference waveform slot (A or B), and 
then recall the reference waveform 
into the display. 
Waveform 
Memory Depth 
The memory depth is limited to 1 M points when 
both channels are activated or 2M points when 
only a single channel is activated. The signal must 
be triggered /stopped to have access to the full 
memory depth. Therefore when a signal is saved 
the waveform will be automatically stopped if it is 
not manually triggered /stopped first. 
There are a number of conditions when all of the 
available memory is not utilized due to a limited 
number of different sample rates. This can be 
caused by an un-triggered signal, or a time/div 
setting that is too fast to display all the points on 
screen. 
 
