Page 122

background image

GDS-1000A Series User Manual

122

4. Press Wavetype repeatedly to

select the standard square
wave.

5. Press the Autoset key. The

compensation signal will
appear in the display.

Autoset

6. Press the Display key, then

Type to select the vector
waveform.

7. Turn the adjustment point on the probe until

the signal edge becomes sharp.

Over

Compensation

Normal

Under

Compensation