beautypg.com

Rear panel – GW Instek GDS-1000A Series User Manual User Manual

Page 18

background image

GDS-1000A Series User Manual

18

Rear Panel

Power cord
socket

Fuse socket

Power cord socket accepts the AC
mains, 100 ~ 240V, 50/60Hz.
The fuse socket holds the AC main
fuse, T1A/250V.
For the fuse replacement
procedure, see page126.

USB slave port

Accepts a type B (slave) male USB
connector for remote control of the
oscilloscope (page89) or to print
directly to a PictBridge compatible
printer.

Calibration
output

Outputs the calibration signal used
in vertical scale accuracy calibration
(page120).

Security lock slot

Standard laptop security lock slot
for ensuring the security of the
GDS-1000A.