beautypg.com

Cursor measurements, Using the horizontal cursors – GW Instek GDS-1000A Series User Manual User Manual

Page 56

background image

GDS-1000A Series User Manual

56

Cursor Measurements

Cursor lines, horizontal or vertical, show the precise position of the
input waveforms or the math operation results. The horizontal
cursors can track time, voltage/current* and frequency, whilst the
vertical cursors can track voltage/current*. All measurements are
updated in real-time. *probe type dependant (page 79).

Using the horizontal cursors

Procedure

1. Press the Cursor key. The

cursors appear in the
display.

Cursor

2. Press X↔Y to select the

horizontal (X1&X2) cursor.

3. Press Source repeatedly to

select the source channel.

Range

CH1, 2, MATH

4. The cursor measurement results will appear in

the menu, F2 to F4.

Parameters X1

Time position of the left cursor. (relative to
zero)

X2

Time position of the right cursor. (relative
to zero)

X1X2

The difference between the X1 and X2.

∆: us The time difference between X1 and X2.

f: Hz

The time difference converted to
frequency.

V/A

The voltage/current difference from X1
and X2.