Yokogawa Button Operated DX2000 User Manual
Page 123
3-62
IM 04L41B01-62EN
Attribute
Set value/choice
(Underlined items are initial
set values)
Description, conditions
Indicator
[SET], [Mark], [Bar]
This is the indicator displayed on the scale plate.
Setting is not possible when “Kind” is [Off].
SET: Setting of the DX recorder
Mark: Present value marks are displayed for the number of channels
registered to the group.
Bar: Bars are displayed for the number of channels registered to the
group.
Indicator type
(can only be set with
a Version 4.01 screen
and a Version 4.11
screen)
[Standard], [Micro], [Small],
[Balance]
Selects the indicator format.
Digit
[SET], [Normal], [Fine]
This is the number of display digits of the scale value displayed on
the scale plate.
Setting is not possible when “Kind” is [Off].
SET: Setting of the DX recorder
Normal: Effective digits of the scale value: 2 digits
Fine: Effective digits of the scale value: 3 digits
Unit
(can only be set with
a Version 4.01 screen
and a Version 4.11
screen)
[Off], [On]
Selects whether or not to display the units.
On: The units are displayed
Off: The units are not displayed
Margin (can only be
set with a Version
4.01 screen and a
Version 4.11 screen)
[Off], [On]
Sets whether or not to display a margin on both sides in the direction
of the span.
On: Margin of 3% of component width is added in the direction of the
span
Off: The margin is not displayed
Display group
switching
(can only be set with
a Version 4.01 screen
and a Version 4.11
screen)
[Off], [On]
Sets whether or not to switch the group display in the DX recorder’s
execution panel when the left/right keys are pressed.
On: Switched
Off: Not switched
Trend direction
[SET], [Vertical], [Horizontal]
This is the waveform flow direction.
SET: Setting of the DX recorder
Vertical: Vertical waveform direction
Horizontal: Horizontal waveform direction
Alarm mark display
[Off], [On], [SET]
Selection of alarm mark display/no display
Setting is not possible when “Kind” is [Off].
Off: No alarm mark display
On: Alarm mark display
SET: Setting of the DX recorder
Alarm mark
[Alarm], [Fixed], [SET]
This selects the alarm mark style.
Setting is not possible when “Kind” is [Off] or “Alarm mark display” is
[Off]
Alarm: Color change may occur at the time of trapezoid display and
at the time of alarm ON.
Fixed: No color change at the time of trapezoid display and at the
time of alarm ON.
SET: Setting of the DX recorder
2nd span
[Off], [On]
This selects effective/disabled for 2nd span.
On: Effective
Off: Disabled
*1
DX1000 max. width = 320, DX2000 max. width = 640
*2
DX1000 min. Y = 24(0), DX2000 min. Y = 40(0), ( ) = Status area set to No display
*3
DX1000 max. height = 216(240), DX2000 max. height = 440(80), ( ) = Status area set to No display
3.29 Attributes of Scale Components