Yokogawa RotaMASS 3-Series User Manual
Page 78

7. pArAMETEr dESCrIpTION
7-6
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
parameter
data
form
data range
dec.
point
Unit
r : read
W : write
Exe :
execute
reset
Master
reset
power on
default
Disp contrast
Select
-5
-
-
R/W
X
0
-4
-3
-2
-1
0
1
2
3
4
5
Disp Period
Select
0.5 s
-
-
R/W
X
1.0 s
1.0 s
2.0 s
Flow direction
Select
Forward
-
-
R/W
X
Forward
Reverse
Mass flow unit
Select
g/s
-
-
R/W
X
RCCS30…33
kg/h
g/min
RCCS34…39/XR
t/h
g/h
kg/s
kg/min
kg/h
kg/d
t/min
t/h
t/d
lb/s
lb/min
lb/h
lb/d
Mass flow format
Select
xxxxxxx
-
-
R/W
X
xxxx.XX
xxxxx.X
xxxx.XX
xxx.XXX
xx.XXXX
x.XXXXX
Mass flow LRV
Decimal
-Qmax to Qmax
0 to 5
Mass flow unit
R/W
X
0
Mass flow URV
Decimal
-Qmax to Qmax
0 to 5
Mass flow unit
R/W
X
Qnom
Mass flow damping
Decimal
0 to 200
1
Sec.
R/W
X
3
Mass flow lowcut
Decimal
0 to 10% of Qmax
0 to 5
Mass flow unit
R/W
X
0
Mass flow alm 1 sel
Select
No function
-
-
R/W
X
No function
More than
Less than
Mass flow alm 1 crit
Decimal
-110% to 110% of
URV
0 to 5
Mass flow unit
R/W
X
0
Mass flow alm 2 sel
Select
No function
-
-
R/W
X
No function
More than
Less than
Mass flow alm 2 crit
Decimal
-110% to 110% of
URV
0 to 5
Mass flow unit
R/W
X
10
Mass flow fix val sel
Select
Inhibit
-
-
R/W
X
Inhibit
Enable
Mass flow fixed val
Decimal
0 to Qnom
0 to 5
Mass flow unit
R/W
X
0