8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-12

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.7.9 Setting "Burn-out" mode

The flowmeter is equipped with a CPU error burn-out function used to set the analog output direction upon
CPU error. When factory-shipment the output of CPU error burn-out is set to HIGH. The setting of the
direction of analog output from burn-out can be changed.

To change the direction of output arising from burn-out, switch the jumper J2 on the CPU Board (see figure
below).

CPU Board

Cable to display

JP2 (Burn out)

High : I ≥ 21.6mA

Low : I ≤ 3.6mA

F810.EPS