7.15 HArT output (detailed Setup)
HArT output
To set the assignments of HART device variables which are reported via
HART Universal Command 3.
pv is
Set the assignment of Primary variable (PV).
PV is linked to Analog output 1. Changing of PV will change the
assignment of Analog output 1 and vice versa.
Select:
Mass flow, volume flow, density, Temperature,
Concentration meas, Net flow, None
Sv is
Set the assignment of Secondary variable (SV).
SV is linked to Analog output 2. Changing of SV will change the
assignment of Analog output 2 and vice versa.
Select:
Mass flow, volume flow, density, Temperature,
Concentration meas, Net flow, None
If Analog output 2 is not available (option /KF2) an extended selection is
available, see “HART Variable list” below.
Tv is
Set the assignment of Tertiary variable (TV).
TV is linked to Pulse output 1 if
pulse/Stat 1 sel is set to pulse. In this
case changing of TV will change the assignment of Pulse output 1 and
vice versa.
Select:
Mass flow, volume flow, density, Temperature,
Concentration meas, Net flow, None, F-Total mass, r-Total mass,
F-Total volume, r-Total volume, F-Total net, r-Total net
If
pulse/Stat 1 sel is set to Status out or No function an extended
selection is available, see “HART Variable list” below . This selection is
stored separately from the selection for “Pulse”.
qv is
Set the assignment of Quaternary variable (QV).
QV is linked to Pulse output 2 if
pulse/Stat 2 sel is set to pulse. In this
case changing of QV will change the assignment of Pulse output 2 and
vice versa.
Select:
Mass flow, volume flow, density, Temperature,
Concentration meas, Net flow, None, F-Total mass, r-Total mass,
F-Total volume, r-Total volume, F-Total net, r-Total net
If Pulse output 2 is not available (option /KF2 ) or
pulse/Stat 2 sel is set to
Status out or No function an extended selection is available, see “HART
Variable list” below.
This selection is stored separately from the selection for “Pulse”.
HArT variable list
(For HArT- communication:
variables which are not
allowed will be rejected by
"Invalid Selection")
Mass flow, volume flow, density, Temperature,
Concentration meas, Net flow, None,
F-Total mass, r-Total mass,
F-Total volume, r-Total volume,
F-Total net, r-Total net,
F-Total flex, r-Total flex,
d-Total mass, d-Total volume,
d-Total net, d-Total flex,
F-Total energy, velocity, drive gain