5. BASIC OpErATINg prOCEdUrES

5-11

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

reading Autozero history (see also chapter 7.24) :

MFR

DEN

TMP

FTL

SET> 2s

Setting Enable?

Main Menu

Language

Process variables

Diag/Service

Main Menu

Diag/Service

Basic Setup

Detailed Setup

SET

2x

Diag/Service

Autozero check

Autozero

Output trim

2x

SET

Autozero check

Autozero history

Autozero initial

Autozero range

Autozero history

05/05/2003

10/06/2003

SET

AZ hist : 05/05/2003

+ 0000.00 kg/h

+ 00.0000 kg/l

+ 0030.00 degC

SET

Diag/Service

F59.EPS

Autozero check

Input/Output test

Selftest/Status

SET

15/07/2004

SHIFT+ SET

Autozero

05/05/2003

10/06/2003

15/07/2004

123.450 kg/h

0.99700 kg/l

12.3400 degC

12.3450 kg

Autozero

Auto zero Exe

XXX.XXkg/h

2x SET

Autozero

Autozero exe

Doing Autozero 180sec

After 30s /180s

Auto zero Exe

Execute

After 3s

Set date with SHIFT+

. Then press 2x SET

Autozero data are written in Autozero
history.

Autozero

Date (DD/MM/YYYY)

00/00/2000

F58a.EPS

3x SHIFT + SET

Measuring mode

Not Execute