8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg
8-1
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
8.1 Error descriptions and countermeasure
The self check function displays instrument errors, warnings and alarms.
When an alarm or error occurs, the code (e.g. E-01) and the name (e.g. Frequency fault) appear on the
display during display mode, alternating with measuring value indication. This does not happen in parameter
setting mode.
The alarms, errors and warnings are written into history, see chapter 9.2.
The following table shows the possible alarms, warnings and error with the concerning influences on outputs
and fields variables.
Alarm (A);
Warning
(W)
Error (E)
Name
Flow on Dis-
play
Density/Temp.
on Display
Totalizer on
Display
Analog
outputs
Frequ./Pulse-
outputs
Status outputs
*)
nc = no change, nd = not defined
A-01
MF 1 criteria
nc
nc
nc
nc
6)
Mass flow alm 1
*)
A-02
MF 2 criteria
nc
nc
nc
nc
6)
Mass flow alm 2
*)
A-03
VF 1 criteria
nc
nc
nc
nc
6)
Vol flow alm 1
*)
A-04
VF 2 criteria
nc
nc
nc
nc
6)
Vol flow alm 2
*)
A-05
Dens. 1 criteria
nc
nc
nc
nc
6)
Density alm 1
*)
A-06
Dens. 2 criteria
nc
nc
nc
nc
6)
Density alm 2
*)
A-07
Temp. 1 criteria
nc
nc
nc
nc
6)
Temp alm 1
*)
A-08
Temp. 2 criteria
nc
nc
nc
nc
6)
Temp alm 2
*)
A-09
Conc. 1 criteria
nc
nc
nc
nc
6)
Conc meas alm 1
*)
A-10
Conc. 2 criteria
nc
nc
nc
nc
6)
Conc meas alm 2
*)
A-11
Netflow 1 crit.
nc
nc
nc
nc
6)
Net flow alm 1
*)
A-12
Netflow 2 crit.
nc
nc
nc
nc
6)
Net flow alm 2
*)
A-13
Total thr exceed
nc
nc
hold;continue
1)
nc
6)
Total switch
*)
A-14
Slug detection
MFL: meas.;
hold
4)
DEN: meas.;
hold
4)
nc
nc
6)
Slug alarm
*)
A-15
Empty pipe det.
MFL: meas.;
hold
2)
nc
nc
nc
6)
Empty pipe alm
*)
A-16
Corrosion det.
nc
nc
nc
nc
6)
Corrosion alm
*)
W-01
Density low
nc
DEN: 0 kg/l
5)
nc
nc
nc
nc
W-02
Fixed dens act.
nc
DEN: fix value
nc
nc
nc
nc
W-03
Fixed temp act.
nc
TEMP: fix value
nc
nc
nc
nc
W-04
Fixed MF active
MFL: fix
value
+ act. zero
nc
nc
nc
nc
nc
W-05
Autozero running
nc
nc
nc
nc
nc
nc
W-06
AZ val error
nc
nc
nc
nc
nc
nc
W-07
AZ fluct error
nc
nc
nc
nc
nc
nc
W-08
Sensor simul.
nc
nc
nc
nc
nc
nc
W-09
Fixed Prim Out
nc
nc
nc
4mA
nc
nc
W-10
Prim Out sat.
nc
nc
nc
nc
nc
nc
W-11
Sec Out sat.
nc
nc
nc
nc
nc
nc
W-12
Tert Out sat.
nc
nc
nc
nc
nc
nc
W-13
Quart Out sat.
nc
nc
nc
nc
nc
nc
E-01
Frequency fault
nd
Dens = nd
Temp = nc
hold
3,6mA;…
3)
6)
7)
E-02
Signal fault
nd
nc
hold
3,6mA;…
3)
6)
7)
E-03
EEPROM fault
nd
nd
hold
3,6mA;…
3)
6)
7)
E-04
CPU fault
nd
nd
nd
3,6mA…
8)
nd
nd
E-05
DSP fault
nd
nd
hold
3,6mA…
8)
6)
7)
E-06
Sensor 1 fault
nd
Dens = nd
Temp = nc
hold
3,6mA;…
3)
6)
7)
E-07
Sensor 2 fault
nd
nc
hold
3,6mA;…
3)
6)
7)
E-08
Temp sens fault
nc
Density = nd
Temp = 30°C
hold
3,6mA;…
3)
6)
7)
1)
Choice of “Total at alarm” (Detailed Setup / Totalizer)
2)
Choice of “Output after Empty pipe” (Detailed Setup / Empty pipe detect)
3)
Choice of “Analog 1/2 alarm out” (Detailed Setup / Config Output) and NAMUR/NOT NAMUR (Measuring range)
4)
Choice of “After Slug” (Detailed Setup / Slug detection)
5)
Only if “Liquid” is selected (Detailed Setup / Liquid/gas selection)
6)
Choice of “Pulse 1 at alarm” or “Pulse 2 at alarm” (Detailed Setup / Pulse/Status out 1/2)
7)
Status outputs active if “All Errors” or “All Alarms & Errors” is selected (Detailed Setup / Pulse/Status out 1/2)
8)
Choice of jumper “Burn-out” (See chapter 8.7.9 Setting “Burn-out” Mode)
*) Status outputs acc. SO 1/2 function and SO 1/2 active mode