8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-1

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.1 Error descriptions and countermeasure

The self check function displays instrument errors, warnings and alarms.

When an alarm or error occurs, the code (e.g. E-01) and the name (e.g. Frequency fault) appear on the

display during display mode, alternating with measuring value indication. This does not happen in parameter

setting mode.

The alarms, errors and warnings are written into history, see chapter 9.2.

The following table shows the possible alarms, warnings and error with the concerning influences on outputs
and fields variables.

Alarm (A);

Warning

(W)

Error (E)

Name

Flow on Dis-

play

Density/Temp.

on Display

Totalizer on

Display

Analog

outputs

Frequ./Pulse-

outputs

Status outputs

*)

nc = no change, nd = not defined

A-01

MF 1 criteria

nc

nc

nc

nc

6)

Mass flow alm 1

*)

A-02

MF 2 criteria

nc

nc

nc

nc

6)

Mass flow alm 2

*)

A-03

VF 1 criteria

nc

nc

nc

nc

6)

Vol flow alm 1

*)

A-04

VF 2 criteria

nc

nc

nc

nc

6)

Vol flow alm 2

*)

A-05

Dens. 1 criteria

nc

nc

nc

nc

6)

Density alm 1

*)

A-06

Dens. 2 criteria

nc

nc

nc

nc

6)

Density alm 2

*)

A-07

Temp. 1 criteria

nc

nc

nc

nc

6)

Temp alm 1

*)

A-08

Temp. 2 criteria

nc

nc

nc

nc

6)

Temp alm 2

*)

A-09

Conc. 1 criteria

nc

nc

nc

nc

6)

Conc meas alm 1

*)

A-10

Conc. 2 criteria

nc

nc

nc

nc

6)

Conc meas alm 2

*)

A-11

Netflow 1 crit.

nc

nc

nc

nc

6)

Net flow alm 1

*)

A-12

Netflow 2 crit.

nc

nc

nc

nc

6)

Net flow alm 2

*)

A-13

Total thr exceed

nc

nc

hold;continue

1)

nc

6)

Total switch

*)

A-14

Slug detection

MFL: meas.;

hold

4)

DEN: meas.;

hold

4)

nc

nc

6)

Slug alarm

*)

A-15

Empty pipe det.

MFL: meas.;

hold

2)

nc

nc

nc

6)

Empty pipe alm

*)

A-16

Corrosion det.

nc

nc

nc

nc

6)

Corrosion alm

*)

W-01

Density low

nc

DEN: 0 kg/l

5)

nc

nc

nc

nc

W-02

Fixed dens act.

nc

DEN: fix value

nc

nc

nc

nc

W-03

Fixed temp act.

nc

TEMP: fix value

nc

nc

nc

nc

W-04

Fixed MF active

MFL: fix

value

+ act. zero

nc

nc

nc

nc

nc

W-05

Autozero running

nc

nc

nc

nc

nc

nc

W-06

AZ val error

nc

nc

nc

nc

nc

nc

W-07

AZ fluct error

nc

nc

nc

nc

nc

nc

W-08

Sensor simul.

nc

nc

nc

nc

nc

nc

W-09

Fixed Prim Out

nc

nc

nc

4mA

nc

nc

W-10

Prim Out sat.

nc

nc

nc

nc

nc

nc

W-11

Sec Out sat.

nc

nc

nc

nc

nc

nc

W-12

Tert Out sat.

nc

nc

nc

nc

nc

nc

W-13

Quart Out sat.

nc

nc

nc

nc

nc

nc

E-01

Frequency fault

nd

Dens = nd

Temp = nc

hold

3,6mA;…

3)

6)

7)

E-02

Signal fault

nd

nc

hold

3,6mA;…

3)

6)

7)

E-03

EEPROM fault

nd

nd

hold

3,6mA;…

3)

6)

7)

E-04

CPU fault

nd

nd

nd

3,6mA…

8)

nd

nd

E-05

DSP fault

nd

nd

hold

3,6mA…

8)

6)

7)

E-06

Sensor 1 fault

nd

Dens = nd

Temp = nc

hold

3,6mA;…

3)

6)

7)

E-07

Sensor 2 fault

nd

nc

hold

3,6mA;…

3)

6)

7)

E-08

Temp sens fault

nc

Density = nd

Temp = 30°C

hold

3,6mA;…

3)

6)

7)

1)

Choice of “Total at alarm” (Detailed Setup / Totalizer)

2)

Choice of “Output after Empty pipe” (Detailed Setup / Empty pipe detect)

3)

Choice of “Analog 1/2 alarm out” (Detailed Setup / Config Output) and NAMUR/NOT NAMUR (Measuring range)

4)

Choice of “After Slug” (Detailed Setup / Slug detection)

5)

Only if “Liquid” is selected (Detailed Setup / Liquid/gas selection)

6)

Choice of “Pulse 1 at alarm” or “Pulse 2 at alarm” (Detailed Setup / Pulse/Status out 1/2)

7)

Status outputs active if “All Errors” or “All Alarms & Errors” is selected (Detailed Setup / Pulse/Status out 1/2)

8)

Choice of jumper “Burn-out” (See chapter 8.7.9 Setting “Burn-out” Mode)

*) Status outputs acc. SO 1/2 function and SO 1/2 active mode

8. SElF-dIAgNOSTIC ANd TrOUBlE-

SHOOTINg