4 unstable zero, 4 unstable zero -7 – Yokogawa RotaMASS 3-Series User Manual

Page 183

background image

8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-7

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.7.4 Unstable zero

no

yes

yes

no

no

yes

no

Is the tube filled with liquid

completely (check the

density)?

Fill the detector. May be

vertical installation is

indicated.

Are bubbles in the liquid ?

Check piping and

installation.

Is the detector installed

free of mechanical stress?

Is electrical grounding

correct according chapter

4

.

8.4; 4.8.5; 9 ?

Make correct grounding.

Is the flowmeter located

near motors, transformers

or other electric

equipments ?

Shield the flowmeter

sufficiently.

Locate the flowmeter as

far as possible from
these equipments.

Is the connection box free

of dirt and humidity? (only

for remote installation)

yes

no

Loose the process

connections and make

sure that they are in line.

Support only one

adjacent pipe, let one

side of the adjacent pipe

be “soft”.

yes

Contact the nearest

YOKOGAWA office or

service center.

yes

no

Clean and dry the

connection box. Make

sure that humidity

cannot penetrate.

F84.EPS