4. INSTAllATION

4-13

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

Rotamass Remote Cable RCCY03 : Converter RCCF31 side

25

7

9

Cut

1

2

3

4

5
6

7

8

9

Shrinkdown

plastic tube

d = 5mm

Shrinkdown

plastic tube

d 3.2mm

Termination procedure:
Cable end detector
- Remove PVC outer sheeting and outer shielding 100 mm from the end.
- Remove the clear wrap and the filler material.
- Remove the foil that is around the isolated wires.
- Clip off each drain wire close to the cable jacket.
- Slide a shrink down plastic tube (d = 3.2 mm, l = 20 mm) over each of the 3 pairs and a shrink down plastic
tube (d = 4.8 mm, l = 20 mm) over the triple of wires, push it to the cable jacket and heat with hot air.
- Strip 8 mm of the cable ends.
- Fix the light blue terminal sleeves (0.25 mm

2

) to the wire endings of the 3 cable pairs.

- Fix the orange terminal sleeves (0.5 mm

2

) to the wire endings of the cable triple

- Make a radial cut into the PVC outer sheeting 25 mm from the end and cut lenghtwise.

Cable end converter
- Remove PVC outer sheeting and outer shielding 100 mm from the end.
- Remove the clear wrap and the filler material.
- Remove the foil that is around the isolated wires.
- Do not clip off the drain wires !
- Twist the 4 drain wires together and fix the 1.5 mm

2

terminal sleeve.

- Slide a shrink-down plastic tube (d = 3.2 mm, l =20 mm) over each of the 3 pairs and a shrink down
plastic tube (d = 4.8 mm, l = 20 mm) over the triple of wires, push it to the cable jacket and heat with hot air.
- Strip 8 mm of the cable ends.
- Fix the light blue terminal sleeves (0.25 mm

2

) to the wire endings of the 3 cable pairs.

- Fix the orange terminal sleeves (0.5 mm

2

) to the wire endings of the cable triple

- Make a radial cut into the PVC outer sheeting 25 mm from the end and cut lenghtwise.
- Slide the conductor markers onto the pairs of wires on both sides of the cable, so that the pairs are
numbered 1-2, 3-4, 5-6 and the triple 7-8-9. Each cable must have the same number on detector and on
converter side.

Rotamass Remote Cable RCCY03 : Detector RCCS side

25

Cut

Terminal sleeves

0.5mm2 (orange)

Terminal sleeves

0.25mm2 (light blue)

Shrinkdown

plastic tube

d 5mm

Shrinkdown

plastic tube

d 3.2mm

1

2

3

4

5
6

7

8

9