7.16 Totalizer functions (Basic or detailed Setup)
Totalizer
To set totalizer function.
Totalizer increments the forward flow if flow direction is forward or
the reverse flow if flow direction is reverse.
Total flex select
Flex totalizer is available as a secondary total for mass, volume or net.
Select:
Mass, volume, Net
If assignment is changed Flex totalizer will be set to zero,
Total flex
unit will be set to default.
Total flex unit
When
Total flex select is set to Mass or Net,
the
Total flex unit g, kg, t, lb can be selected.
When
Total flex select is set to volume or Net,
the
Total flex unit cm³, l, m³, gal, kgal, Cuft, bbl, Impgal,
kImpgal, l(N), m³(N), Sl, Scuft, MMscuft, Sm³ can be selected.
l(N), m³(N) only for gas application.
Sl, Scuft, MMscuft, Sm³ only for gas application or if Reference
density value is selected.
Total mass unit
Select:
g, kg, t, lb
Total volume unit
For
Total volume unit
cm³, l, m³, gal, kgal, Cuft, bbl, Impgal, kImpgal, l(N), m³(N), Sl,
Scuft, MMscuft, Sm³
can be selected.
l(N), m³(N) only for gas application.
Sl, Scuft, MMscuft, Sm³ only for gas application or if reference
density value is selected.
Total net unit
For option /CST:
When
Net flow select is set to Mass,
the
Total net unit g, kg, t, lb can be selected.
When
Net flow select is set to volume,
the
Total net unit cm³, l, m³, gal, kgal, Cuft, bbl, Impgal, kImpgal
can be selected.
For option /Cxx:
Selection depends from concentration unit.
Total energy unit
Select:
MJ, Btu
Energy factor/Heat of combustion Set the calorific value of the medium with the unit Energy factor/
Heat of combust. unit.
The totalized energy of the medium can be indicated on display by
F-Total energy.
Heat of combust. unit
MJ/kg, Btu/lb, MJ/m³(N) and Btu/Scuft can
be selected.
Total switch select
Select:
Flex Total, Mass Total, volume Total, Net Total
Total limit switch
A totalizer limit can be set only to the forward
totalizer selected in
Total select. If status output is set to Total
switch, status output is set active, if the totalizer reaches this value.
Total switch is not active if set to 0.0.
Total at alarm
Select:
Hold, Continue Total
Characteristic of totalizer selected in
Total select in case of the
Total switch is active.
If
Hold is selected forward and reverse totalising stops.
Total reset cntrl
Select:
Inhibit, Enable
Enables or inhibits total reset.