5 setting parameters in converter with option /nc – Yokogawa RotaMASS 3-Series User Manual
Page 60

5. BASIC OpErATINg prOCEdUrES
5-8
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
5.5 Setting parameters in converter with option /NC
If remote converter RCCF31 / RCCR31 was ordered with option /NC (no combination), the customer must
set the parameters of the connected detector by himself.
In parameter Detailed Setup / Sensor model the sensor model must be selected. Then, the sensor model
depending parameters are set automatically to default value :
Qnom, Mass flow unit, Mass flow LRV, Mass flow URV, Vol flow unit, Vol flow LRV, Vol flow URV, Autozero
range, Autozero fluctuation range.
In parameter Detailed Setup / Sensor constants the sensor constants from calibration certificate must be set.
If material of wetted parts is HC see instruction TI 01R04B04-03E-E attached to the converter.
In Parameter Basic setup / Tag or Basic setup / Long tag the tag number must be set if necessary.