7. pArAMETEr dESCrIpTION

7-101

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

7.21 Empty pipe detection (detailed Setup)

Empty pipe detection

Not available if gas measurement
(option /GA) is ordered

Detection of empty pipe

After

Empty pipe crit has been reached, alarm

output is activated if selected.
This function can also be used to detect tube failure
(see chapter 8.8).

Empty pipe alm sel

Select:

Not apply, Apply

Empty pipe crit

Set density value as criteria when empty pipe alarm
should happen. If

Empty pipe crit is smaller than

density Urv (or LRV, whichever is higher), alarm
appears if density is lower than

Empty pipe crit.

If

Empty pipe crit is equal or greater than density

Urv (or LRV, whichever is higher), alarm appears if
density is higher than

Empty pipe crit.

After empty pipe

Select the behaviour of the outputs after empty pipe
has happened.

Select:

Massflow = Zero, Measured value, Hold

With this function the outputs can be set to zero if the
detector runs empty during maintenance. It is also
useful for batching if the batch starts with an empty
pipe. In such case please install the detector vertically.

Gas application?

Yes

No

Empty pipe alm sel = Apply?

Yes

No

Empty pipe crit < Density URV?

(or LRV, whichever is higher)

Empty pipe crit ≥ Density URV?

(or LRV, whichever is higher)

Yes

No

Yes

No

Empty pipe OFF

Alarm off

Empty pipe

ON

Alarm On

if density <

empty pipe crit

Outputs either

Measured

Value, Hold or

4mA/0Hz

Empty pipe

OFF

Alarm off

Empty pipe

ON

Alarm On

if density ≥

empty pipe crit

Outputs either

Measured

Value, Hold or

4mA/0Hz

Empty pipe

OFF

Alarm off

Empty pipe

OFF

Alarm off