8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-11

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.7.8 discrepancy of output signals to the assigned measurand

no

yes

Measured temperature

values OK ?

Converter fault.

Contact the nearest

YOKOGAWA office or

service center.

Check the electronics:

Connect Pt100 simulator to terminals TP1, TP2, TP3 in converter

RCCF31 terminal box.

TP1 : + power, + sensor

TP2 : - sensor

TP3 :

Check your reference

temperature.

F88.EPS

- power

no

yes

no

yes

Is the setting of the

doubtful output signal

correct ?

Make correct parameter

setting.

Check output according

chapter 8.4.

Output OK ?

Contact the nearest

YOKOGAWA office or

service center

Check measurand.

F89.EPS