9.2.2 Installation
Integral type RCCT3 :
L/+
N/-
G
Power supply
I/O control
Hazardous area
Safe area
I/O
Rotamass RCCT3
F91.EPS
Terminal Box
For customer insulation of RCCS30 to 39/XR the following must be regarded :
The table "with factory insulation" is calculated with 80 mm insulation and k-factor = 0.4 W/m
2
K.
If your insulation data are worse than these use table "without insulation" !
NOTE
RCCS30 to RCCS33 without
insulation
RCCS30 to RCCS33
with factory insulation
Temp.
class
Max. ambient
temperature
Max. process
temperature
Max. ambient
temperature
Max. process
temperature
T6
50°C / 122°F
60°C / 140°F
60°C / 140°F
60°C / 140°F
T5
50°C / 122°F
80°C / 176°F
80°C / 176°F
90°C / 194°F
T4
80°C / 176°F
50°C / 122°F
100°C / 212°F
120°C / 248°F
80°C / 176°F
130°C / 266°F
T3
80°C / 176°F
150°C / 302°F
80°C / 176°F
150°C / 302°F
T2
80°C / 176°F
150°C / 302°F
80°C / 176°F
150°C / 302°F
RCCS34 to RCCS39/XR
without insulation
RCCS34 to RCCS39/XR
with factory insulation
RCCT34 to RCCT39/XR
Temp.
class
Max. ambient
temperature
Max. process
temperature
Max. ambient
temperature
Max. process
temperature
Max. ambient
temperature
Max. process
temperature
T6
40°C / 104°F
40°C / 104°F
65°C / 149°C
65°C / 149°F
50°C / 122°F
65°C / 149°C
T5
55°C / 131°F
55°C / 131°F
75°C / 167°F
75°C / 167°F
50°C / 122°F
80°C / 176°F
T4
80°C / 176°F
40°C / 104°F
100°C / 212°F
120°C / 248°F
70°C / 158°F
115°C / 239°F
50°C / 122°F
115°C / 239°F
T3
80°C / 176°F
40°C / 104°F
160°C / 320°F
180°C / 356°F
70°C / 158°F
180°C / 356°F
50°C / 122°F
150°C / 302°F
T2
80°C / 176°F
220°C / 428°F
65°C /149°F
275°C / 527°F
50°C / 122°F
150°C / 302°F
T1
80°C / 176°F
220°C / 428°F
45°C / 113°F
350°C / 662°F
50°C / 122°F
150°C / 302°F