4. INSTAllATION
4-15
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
4.8.5 power supply wiring
The terminals for the power supply and for the I/O ports are in a separate terminal box as shown in the
figures below.
Cable connection to rack-mount converter RCCR31:
1.
Loose the cable clamp on terminal board
2.
Remove the 25 mm section of PVC outer sheathing from the cable and fix the outer ring of the cable
with the cable clamp on terminal board. (see left figure below).
3.
Connect the numbered leads to the terminals as shown in the right figure below.
4.
Connect inner shields to terminal COM.
Detector
Cable
Converter
RCCS3
RCCY03
RCCR31
D+
1
1
D+
D-
2
2
D-
S1+
3
3
S1+
S1-
4
4
S1-
S2+
5
5
S2+
S2-
6
6
S2-
TP1
7
7
TP1
TP2
8
8
TP2
TP3
9
9
TP3
Shields
COM
'BJ
$PWFSMPDLJOHTDSFXT
$PWFSMPDLJOHTDSFXT
%JTQMBZ
%JTQMBZ
3$$5*OUFHSBMUZQF
3$$'SFNPUFDPOWFSUFS
1SPUFDUJWFFBSUI
UFSNJOBM .TDSFX
1PXFSUFSNJOBMCMPDL
*OQVU0VUQVUTJHOBM
UFSNJOBMCMPDL