7. pArAMETEr dESCrIpTION

7-85

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

7.7 density functions (Basic or detailed Setup)

density

To set the functions of density measurement.
Notice: Density measurement is only reliable if there is no gas in the
liquid. For liquid measurement a density lower 0.3 kg/l is displayed as
0.0 kg/l. For gas measurement set a fixed density value.

density unit

Select:
g/ml, kg/l, kg/m³, lb/gal, lb/Cuft, g/cm³, g/l, °Bé hv, °Bé lt, °ApI

°Bé hv, °Bé lt, °ApI only with option /CST

density format

Select:

xxxxxxx, xxxxx.X, xxxx.XX, xxx.XXX, xx.XXXX, x.XXXXX

density lrv
(low range value)

Set a density value to the 4 mA or 0 Hz.

density Urv
(upper range value)

Set a density value to the 20 mA or xxxxx Hz.
Abs(URV-LRV) ≥ 0.05 kg/l

density damping

damping

Set the damping time in s. Damping time affects outputs and display.
After 5 times damping time the given value is displayed (t=63%)

density lowcut

lowcut

Set the low cut. Low cut affects outputs and display.

density alm 1 sel

alm 1 sel

Select:

No function, More than, less than

density alm 1 crit

alm 1 crit

Select a density alarm level. What happens in case of an alarm has to
be defined in Status out (7.12, 7.13).

density alm 2 sel

alm 2 sel

Select:

No function, More than, less than

density alm 2 crit

alm 2 crit

Restrictions:
All alarms have to be in 0-110%
of URV-LRV
Alarm criteria 2 must always be
larger than alarm criteria 1.

Select a density alarm level. What happens in case of an alarm has to
be defined in Status out (7.12, 7.13).

density fix val sel

Select:

Inhibit, Enable

reference density

Select a density value. For gas measurement (option /GA) set here
your actual density for output of volume flow. For output of standard or
normal volume flow set the standard or normal density of your gas
accordingly. For liquid measurement this parameter can also be used to
enter the standard density of your liquid for output of standard volume
flow or for the actual density in case of density disturbances.

density offset

If the Rotamass detector is installed horizontally (bend down) the value
from the following table must be entered according to the concerning
sensor model.

RCCS30

13.50 g/l

RCCS31

1.80 g/l

RCCS32

2.00 g/l

RCCS33

0.60 g/l

RCCx34

0.90 g/l

RCCx36

0.35 g/l

RCCx38 to 39/XR

0 g/l (default)

pressure

Set the average static line pressure.
Static pressure effect on density and mass flow will be corrected auto-
matically.

pressure unit

Select:

bar, psi, Mpa

ROTAMASS is density-calibrated vertically. If the ROTAMASS detector is installed horizontal an density
offset value from the above table has to be entered to the parameter reference density. As positive values if
the tubes are down and as negative value if the tubes are up..