Yokogawa RotaMASS 3-Series User Manual
Page 166

7. pArAMETEr dESCrIpTION
7-94
IM 01R04B04-00E-E 8th edition March 01, 2011 -00
All Rights Reserved. Copyright © 2003, Rota Yokogawa
Detailed information about using the pulse and status output see chapter 7.12.
SO 2 function
Select:
No Function, Flow direction, Total limit switch, Mass flow
alm 1, Mass flow alm 2, Mass flow alm 1+2, vol flow alm 1, vol flow
alm 2, vol flow alm 1+2, density alm 1,density alm 2, density alm
1+2, Temp alm 1, Temp alm 2, Temp alm 1+2, Concentration meas
alm 1, Concentration meas alm 2, Concentration meas alm 1+2,
Net flow alm 1, Net flow alm 2, Net flow alm 1+2, Slug alarm, Empty
alarm, Corrosion alarm, All alarms, All errors, All alarms & errors
If
Flow direction is selected, status output becomes active for forward
flow and not active for reverse flow, if parameter
flow direction set to
forward. It is the other way round, if parameter flow direction is set to
reverse (see 7.17).
If
Total limit switch is selected, status output becomes active, when
the totalizer reaches the value set to parameter
Total limit switch (see
7.17).
If an alarm function is selected, status output becomes active, if the
concerning alarm occurs.
If
All alarms is selected, status output becomes active, if at least one
alarm occurs. The same applies for errors with the setting All errors.
For
All alarms & errors the status output becomes active, if at least
one alarm or one error occurs.
SO2 active mode
Select:
On Active, Off Active
The output levels depending on this parameter are shown on page 7-91.
7.14 Status input functions (Basic or detailed Setup)
Status input
To set the
status input.
SI function
Select:
No function, Autozero
reset flex totals, reset mass totals, reset volume totals,
reset net totals, reset all totals, 0% signal lock
If
Autozero is selected, active status input starts autozero procedure.
If parameter
Total reset cntrl is enabled:
-
If
reset flex totals is selected, active status input sets flex
totalizer to zero.
-
If
reset mass totals is selected, active status input sets mass
totalizer to zero.
-
If
reset volume totals is selected, active status input sets
volume totalizer to zero.
-
If
reset net totals is selected, active status input sets
net totalizer to zero.
-
If
reset all totals is selected, active status input sets all
totalizer to zero.
SI active mode
Select:
On Active, Off Active
If
On Active is selected, the status input is closed active.
If
Off Active is selected, the status input is open active.