8. SElF-dIAgNOSTIC ANd TrOUBlESHOOTINg

8-3

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

8.2 reading Event + Error History (diag/Service, Self

test/Status)

Event Overview/
Error
Alarm
Warning

Reads the actual errors, alarms and warnings (EAW). In Event
Overview the number of the actual EAWs is shown. In

Error, Alarm

or

Warning the description of the actual events can be found.

Hist Overview in Order

Shows the latest 10 EAWs. By comparing this with

Hist Overview

abs it can be checked if certain EAWs have vanished.

Hist Overview in Order/
Clear History

Select:

Execute, Non execute

If

Execute is selected the latest 10 EAWs are cleared.

Hist Overview abs/
Error
Alarm
Warning

In

Hist Overview abs the number of EAWs after the last clearance

can be viewed. In

Error, Alarm or Warning the description of the

events after the last clearance can be found.

Hist Overview abs/
Clear History

Select:

Execute, Not Execute

If

Execute is selected the user is asked to fill in a date (mm/dd/yyyy).

If a date is set the errors and warnings are stored in the

Error history

with this date.

Error History

In

Error History the number of EAWs of the last clearance of the

Hist Overview abs can be viewed. In Error or Warning the
description of the events can be found. The last 10 entries can be
selected by date.

8.3 Self test (diag/Service)

Self test/ Status

Select:

Self test, Status

If

Self test is selected, all error and alarm status are checked.

If

Status is selected, the existing errors and alarms are indicated in

succession.

8.4 Signal- and I/O-Test (diag/Service)

Input/Output test

These functions enable you to simulate the I/O signals (loop test).
The outputs can be checked with a multimeter.
Refer to the connecting diagrams in chapte

r 4.8.6

. The test can be

cancelled by pressing any key.

Analog output 1
Analog output 2
pulse output 1
Status output 1
pulse output 2
Status output 2
Status input

Setting range: 2 to 22 mA
Setting range: 2 to 22 mA
Setting range: 0 to 10000 Hz
Setting range: On Active or Off Active
Setting range: 0 to 2000 Hz
Setting range: On Active or Off Active
Indication range: Open or Short