beautypg.com

Yokogawa RotaMASS 3-Series User Manual

Page 75

background image

7. pArAMETEr dESCrIpTION

7-3

IM 01R04B04-00E-E 8th edition March 01, 2011 -00

All Rights Reserved. Copyright © 2003, Rota Yokogawa

parameter

data
form

data range

dec.

point

Unit

r : read
W : write
Exe :
execute

reset

Master

reset

power on

default

Analog output 1 mA

Decimal

-

2

mA

R

-

-

Analog output 2

Decimal

-

as sel.
Var.

as sel. Var.

R

-

-

Analog output 2%

Decimal

-

2

%

R

-

-

Analog output 2 mA

Decimal

-

2

mA

R

-

-

Pulse output 1

Decimal

-

1

Hz

R

-

-

Status output 1

ASCII

-

-

-

R

-

-

Pulse output 2

Decimal

-

1

Hz

R

-

-

Status output 2

ASCII

-

-

-

R

-

-

View input

ASCII

-

-

-

R

-

-

Self Test

Select

Not Execute

-

-

R/W

X

Not Execute

Execute

Status

ASCII

Results of last self
test

-

-

R

-

-

Hist Overview ord

ASCII

-

-

-

R

-

-

Hist Overview abs

ASCII

-

-

-

R

-

-

Error History

ASCII

-

-

-

R

-

-

Analog output 1 (Test)

Decimal

2.0 to 22.0

2

mA

W

X

X

4

Analog output 2 (Test)

Decimal

2.0 to 22.0

2

mA

W

X

X

4

Pulse output 1 (Test)

Decimal

0 to 10000

1

Hz

W

X

X

1000

Status Output 1 (Test)

Select

On Active

-

-

W

X

X

On Active

Off Active

Pulse output 2 (Test)

Decimal

0 to 10000

1

Hz

W

X

X

1000

Status Output 2 (Test)

Select

On Active

-

-

W

X

X

On Active

Off Active

Status input (Test)

ASCII

Open/Short

-

-

R

-

-

Autozero tuning

Select

Inhibit

-

-

R/W

X

X

Inhibit

Enable

Autozero Time

Select

3 min

-

-

R/W

X

X

3 min

30 s

Autozero Exe

Select

Not Execute

-

-

Exe

X

X

Not Execute

Execute

Autozero Value

Decimal

depends on size

3

kg/h

R

X

0

Autozero History

ASCII

-

-

-

R

X

-

-

Autozero initial

Decimal

depends on size

3

kg/h

R

X

-

-

Autozero range

Decimal

depends on size

3

kg/h

R

-

-

Autozero fluctuation range

Decimal

depends on size

3

kg/h

R

X

-

-

Output Trim AO 1 trim 4mA Decimal

ACC To HART

-

-

R/W

-

0

Output Trim AO 2 trim 4mA Decimal

ACC To HART

-

-

R/W

-

0

Output Trim AO 1 trim
20mA

Decimal

ACC To HART

-

-

R/W

-

0

Output Trim AO 2 trim
20mA

Decimal

ACC To HART

-

-

R/W

-

0

Tag

ASCII

8 characters

-

-

R/W

Space

Long Tag

ASCII

22 characters

-

-

R/W

Space