<8. Operation of DYFMVTool>
8-23
IM 01C25R51-01E
Table 8.12
Input conditions in Connected Device’s AI Block
Input
Device
Function
Block
Parameter
Value
Unit
Volumetric Flow
Rate
digitalYEWFLO
AI3 Block OUT
or
AI3 Block
SIMULATE
Output.Value
(Relative Index: 8)
or
Simulate.Simulate Value
(Relative Index: 9)
Output Scale.Units
Index
(Relative Index: 11)
Temperature
digitalYEWFLO
(In case
Using built-in
temperature
output)
AI2 Block OUT
or
AI2 Block
SIMULATE
Output.Value
(Relative Index: 8)
or
Simulate.Simulate Value
(Relative Index: 9)
Output Scale.Units
Index
(Relative Index: 11)
Temperature
Transmitter
(In case Using
external
temperature
output)
AI Block OUT
or
AI Block
SIMULATE
Output.Value
(Relative Index: 8)
or
Simulate.Simulate Value
(Relative Index: 9)
Output Scale.Units
Index
(Relative Index: 11)
Pressure
Pressure
Transmitter
AI Block OUT
or
AI Block
SIMULATE
Output.Value
(Relative Index: 8)
or
Simulate.Simulate Value
(Relative Index: 9)
Output Scale.Units
Index
(Relative Index: 11)
IMPORTANT
Before write parameter, set the Function Block Mode in AI Block according to Table 8.12