2 fluid type setup, Fluid type setup -17 – Yokogawa EJX930A User Manual
Page 133

<8. Operation of DYFMVTool>
8-17
IM 01C25R51-01E
8.2.2 Fluid Type Setup
Set parameters (1) to (9) according to Arithmetic Type.
Figure 8.17 Fluid Type Setup Screen
(1), (2) Base Temperature
Set the standard temperature and select temperature unit.
(3), (4) Base Pressure abs
Set the standard abs pressure and select pressure unit.
(5) Deviation
Set the gas deviation factor.
(6), (7) Base Density
Set the standard density and select density unit.
(8) First Temp. Coef.
Set 1
st
temperature factor for density compensation equation in liquid
(9) Second Temp. Coef.
Set 2
nd
temperature factor for density compensation equation in liquid