beautypg.com

Current process input procedure, Input x (1 to 3) setup and calibration, Process output procedure – Watlow Series F4P User Manual

Page 25: Output 1a setup and calibration, Output 1b setup and calibration

background image

Wa t l o w S e r i e s F 4 P

F a c t o r y P a g e

5 . 5

4. Enter 10.000V from the voltage source to the

controller. Allow at least 10 seconds to stabilize.
Press

. once at the Calibrate Input 1 prompt. At

the 10.000V prompt press

. once and to store the

10.000V input press

> once.

5. Rewire for operation and verify calibration.

Current Process Input Procedure

Equipment

• Precision current source, 0 to 20mA range, with

0.01mA resolution.

Input x (1 to 3) Setup and Calibration

Inputs 2 and 3 appear only if the controller is the
enhanced version (F4P _ - _ _ AB - _ _ _ _).

1. Connect the correct power supply to terminals 1, 2

and 3 (see the Wiring Chapter and the Appendix).

2. Connect the current source to input 1 terminals 60

(+) and 62 (-), input 2 terminals 54 (+) and 58 (-), or
input 3 terminals 52 (+) and 56 (-).

3. Enter 4.000mA from the current source to the

controller. Allow at least 10 seconds to stabilize.
Press the Right Key

. once at the Calibrate Input 1

prompt. At the 4.000mA prompt press

. once and to

store 4.000mA press the Up Key

> once.

4. Enter 20.000mA from the current source to the

controller. Allow at least 10 seconds to stabilize.
Press

. once at the Calibrate Input 1 prompt. At

the 20.000mA prompt press

. once and to store

20.000mA press

> once.

5. Rewire for operation and verify calibration.

Process Output Procedure

Equipment

• Precision volt/ammeter with 3.5-digit resolution.

Output 1A Setup and Calibration

1. Connect the correct power supply to terminals 1, 2

and 3 (see the Wiring Chapter and the Appendix).

Milliamperes

2. Connect the volt/ammeter to terminals 42 (+) and 43

(-).

3. Press the Right Key

. at the Calibrate Output 1A

prompt. At the 4.000mA prompt press

. once. Use

the Up Key

> or the Down Key < to adjust the

display to the reading on the volt/ammeter. The
controller should stabilize within one second. Repeat
until the volt/ammeter reads 4.000mA. Press

. to

store the value.

4. Press the Right Key

. at the Calibrate Output 1A

prompt. At the 20.000mA prompt press

. once. Use

the Up Key

> or the Down Key < to adjust the

display to the reading on the volt/ammeter. The
controller should stabilize within one second. Repeat
until the volt/ammeter reads 20.000mA. Press

. to

store the value.

Volts

5. Connect the volt/ammeter to terminals 44 (+) and 43

(-).

6. Press the Right Key

. at the Calibrate Output 1A

prompt. At the 1.000V prompt press

. once. Use the

Up Key

> or the Down Key < to adjust the display

to the reading on the volt/ammeter. The controller
should stabilize within one second. Repeat until the
volt/ammeter reads 1.000V. Press

. to store the

value.

7. Press the Right Key

. at the Calibrate Output 1A

prompt. At the 10.000V prompt press

. once. Use

the Up Key

> or the Down Key < to adjust the

display to the reading on the volt/ammeter. The
controller should stabilize within one second. Repeat
until the volt/ammeter reads 10.000V. Press

. to

store the value.

8. Rewire for operation and verify calibration.

Output 1B Setup and Calibration

1. Connect the correct power supply to terminals 1, 2

and 3 (see the Wiring Chapter and the Appendix).

Milliamperes

2. Connect the volt/ammeter to terminals 39 (+) and 40

(-).

3. Press the Right Key

. at the Calibrate Output 1B

prompt. At the 4.000mA prompt press

. once. Use

the Up Key

> or the Down Key < to adjust the

display to the reading on the volt/ammeter. The
controller should stabilize within one second. Repeat
until the volt/ammeter reads 4.000mA. Press

. to

store the value.

4. Press the Right Key

. at the Calibrate Output 1B

prompt. At the 20.000mA prompt press

. once. Use

the Up Key

> or the Down Key < to adjust the

display to the reading on the volt/ammeter. The
controller should stabilize within one second. Repeat
until the volt/ammeter reads 20.000mA. Press

. to

store the value.

Volts

5. Connect the volt/ammeter to terminals 41 (+) and 40

(-).

6. Press the Right Key

. at the Calibrate Output 1B

prompt. At the 1.000V prompt press

. once. Use the

Up Key

> or the Down Key < to adjust the display

to the reading on the volt/ammeter. The controller
should stabilize within one second. Repeat until the
volt/ammeter reads 1.000V. Press

. to store the

value.