D
Accessories
Calibrating Accessories
If you are installing accessories along with your original system installation, calibrate
your system with a one-time Instrument Calibration wizard. Complete instructions
are in the installation guide that arrives with your imager. If you acquire new
conversion screens, light sources, or filters after your original system installation,
you will have to recalibrate your imager to use them.
Installing Optional Accessories
UV/White Light Conversion Screen
This optional white light conversion screen (catalog #170-8289) converts the UV
light generated in the universal hood to white light. Your imager must be calibrated
to use the white light conversion screen.
To use the white light conversion screen
1.
Center the conversion screen on the imager stage.
2.
Center your samples on top of the conversion screen.
3.
Image the gel using your preferred application