Displaying Gel Images

User Guide

| 101

The Low progress indicator determines which intensity value is shown at

the minimum gray scale (or other color) in the gel image.

The Gamma progress indicator changes the gray scale curve. A value of 1

is linear. A value <1 redistributes a greater proportion of the gray scale to

the first half of the intensity values. A value >1 redistributes a greater

proportion of the gray scale to the second half of the intensity values.

You can also type numerical values in the boxes next to the progress indicators.

Clicking anywhere on the progress indicator bars moves the progress indicator

incrementally.

Options:

Invert image display — inverts dark bands on a light background to light

bands on a dark background. Light bands on a dark background are

inverted to dark bands on a light background.

Highlight saturated pixels — when this checkbox is selected, areas of the

image with saturated signal intensity (higher than a measurable range) are

highlighted in red.

Linear or logarithmic histogram — this adjustment changes the y-axis on

the histogram to display the number of pixels at each intensity value using

either a linear or a logarithmic scale.