Fuse Replacement

Caution:

Always unplug the instrument before changing or checking the fuses.

Power outlet

Fuse holder

This unit is protected by two fuses (5 x 20 mm, 2A Slo-Blo). The fuses are located in

the left side rear of the universal hood, in a fuse holder housed in the power entry

module.

To replace the fuses

1.

Unplug the main power cable from the power outlet.

2.

Use a screwdriver as a lever on the fuse holder to extract the fuse.

3.

Remove the blown fuses and replace them with two new fuses
(catalog

#900-8935).

4.

Slide the fuse holder into the power entry module until it snaps into place.