Fuse Replacement
Caution:
Always unplug the instrument before changing or checking the fuses.
Power outlet
Fuse holder
This unit is protected by two fuses (5 x 20 mm, 2A Slo-Blo). The fuses are located in
the left side rear of the universal hood, in a fuse holder housed in the power entry
module.
To replace the fuses
1.
Unplug the main power cable from the power outlet.
2.
Use a screwdriver as a lever on the fuse holder to extract the fuse.
3.
Remove the blown fuses and replace them with two new fuses
(catalog
#900-8935).
4.
Slide the fuse holder into the power entry module until it snaps into place.