Metrohm NIRS XDS Process Analyzer – DirectLight/NonContact User Manual

Page 85

background image

▪▪▪▪▪▪▪

83

Be sure the fiber probe end is securely installed into the barrel (outer probe part) and proceed using
this set of instructions.

From the Diagnostics menu, select Create Window
Correction.

NOTE: Window Correction must be enabled under
Configure, Options. When enabled, a new DCM
must be created for Window Correction to take
effect.

Vision requests a standard file. Select the file
shown, from the C:\Vision directory. This file
corrects the fiber optic path for the effects of the
barrel (window) being installed.

The TSSZERO file is shown as a spectral plot. The
spectrum is a flat line at 0 AU.

Vision takes a reference using the fiber optic
probe. This spectrum is plotted (not shown here.)

When this message appears, click on “OK”.