Metrohm NIRS XDS Process Analyzer – DirectLight/NonContact User Manual

Page 44

background image

42

▪▪▪▪▪▪▪

11.

Route AC Mains wiring into the housing, and in

to the Main Power Wiring Header as shown at
right.

Terminals 3 and 6 should be left as is, with the
black and white wires attached.

Use table below for wire color codes.

L1, Neutral

L2, HOT

Ground

Intl: Blue

US: White

Intl: Brown

US: Black

Intl: Green/Yellow

US: Green

12.

Attach Grounds to the Ground Lug inside the Base Unit of the EPCU.

13.

The wiring diagram on page 13 of the BEBCO Industries Series 4000 manual is shown for

reference. Refer to the actual diagram for complete wiring particulars.

Follow the actual wiring diagram in the BEBCO manual.

14.

Verify that all wires are secure and that the clamping screws are tight.