8 rj-45 network connection, 1 basic process instrument rj-45 connection, Rj-45 network connection – Metrohm NIRS XDS Process Analyzer – DirectLight/NonContact User Manual

Page 47: Basic process instrument rj-45 connection, 8rj-45 network connection

background image

▪▪▪▪▪▪▪

45

8

RJ-45 Network Connection

RJ-45 Cable Connection applies to the non-purged instrument, and is detailed below. For the purged
unit, fiber optic modems are provided, to meet hazardous location requirements.

8.1

Basic Process Instrument RJ-45 Connection

The Basic XDS Process Analytics Instrument (non-Purged)
has an RJ-45 cable entrance on the left side of the box,
toward the bottom. It is shown, with cable connected.

The plug shown is a water-tight, dust-tight connector
with O-ring seals. It is suitable for NEMA 4X
environments.

A plug kit is provided. Assembly instructions follow.

The AC cable shown is for temporary use only, during
initial acceptance testing. It must be replaced by proper
AC Mains wiring when permanent installation is
performed.

The RJ-45 cable termination kit is shown at right. Verify
that all parts are present. The RJ-45 cable should be
prepared for installation of the termination kit. Approved
methods should be used. If in doubt, instructions are
available on the Belden web site at:

http://bwcecom.belden.com/college/installs/rj45term.htm

Slide the large knurled collar, then the small hex collar,
onto the cable end.

Install the O-ring onto the cable connector housing as
shown. Do not damage or distort the O-ring during
installation.