23

3.3

Weight

System Type

Pounds

Kilograms

Basic System:

93 lbs

41.9

With Vortec Cooler:

106 lbs

47.7

With Purge Control and Air Conditioning:

135 lbs

60.8

3.4

Operating Temperature Range

System Type

Fahrenheit

Centigrade

Basic System:

32-95

0-35

With Vortec Cooler:

32-104

0-40

With Purge Control and Air Conditioning:

32-125

0-51

With Air Conditioning

32-125

0-51

3.5

Operating Humidity Range

System Type

% Relative Humidity

All Systems:

10-90 % non-condensing

3.6

Power Requirements

AC Mains Power: 100/240 VAC Selectable, 50/60 Hz, 2A / 1.5A

Power Consumption: 100W Typical

Heater adds up to 150W

Fans add 20W

With Air Conditioning (optional) 5A / 2.7A