1.

Loosen the 1/16” set screw that secures the

round green housing to the base of the
Electronic Power Control Unit (EPCU). (It is not
tightened at the factory, and may not need to
be loosened.)

Rotate the green cast cover counter-clockwise
until it can be removed from the square base.
This takes about 8-10 turns.

2.

When the green cast cover is removed, the

Control Module Assembly will be exposed as
shown.

Using a #1 Phillips screwdriver, remove both
screws holding the Display Interface Module
face in place.

3.

Gently lift the Display Interface Module face

plate off. Put in safe place.

4.

Unplug both wired headers, as shown, from the

Encapsulated Intrinsic Safety Module (EISM).