27

The vertical distance between the center lines of the upper and lower mounting feet is 24”
(609.6mm).

Leave adequate mounting space for all installed options, as well as room for air circulation and wiring
access. Recommended minimum access and mounting space is as follows:

SYSTEM

Left Side

Right Side

Top

Bottom

Front Access

Basic Unit,
Fan-cooled

12” (305mm)

18” (457mm)

18” (457mm)

Flush Mount

48” (1220mm)

Vortec

Cooled

12” (305mm)

18” (457mm)

18” (457mm)

Flush Mount

48” (1220mm)

Purged
System,
Vortec Cooled

12” (305mm)

18” (457mm)

18” (457mm)

Flush Mount

48” (1220mm)

Purged
System, Air
Conditioned

12” (305mm)

18” (457mm)

18” (457mm)

Flush Mount

48” (1220mm)

The mounting bolts must be 3/8” (10mm) in size. The environment may require stainless steel
mounting hardware due to industry or washdown requirements.

See section 5.4 for more specific information on lifting, handling and mounting the enclosure.

Probe Mounting Information:

Because probes come in various configurations, we have included drawings of the standard probe
designs in section 14, Appendix. Please note the following information about probes:

1.

Probe designs shown are not necessarily the most current revision, due to printing deadlines

and other factors

2.

Probe lengths and materials may be different from designs shown.

3.

Setscrews, mounting dimensions and style are subject to change without notice.

Always request a current drawing (or verify specifications) before committing to mounting
dimensions.