background image

32

▪▪▪▪▪▪▪

RJ-45 Communications fibers enter
the purged system where shown in
the photo, near the filter unit.

The instrument air input to the filter
is shown.

AC Power enters the purged
enclosure via the cable entrance at
lower right in this photo. The XDS
Process Analytics instrument is
shipped with a temporary AC power
cord, to be used for initial
qualification testing only.

Upon permanent installation, the AC
cord is to be removed, and replaced
with permanent wiring that meets
local codes.

5.4

Mounting of XDS Process Enclosure

These steps explain physical mounting of the XDS Process Enclosure. Note that the enclosure may
weigh as much as 135 pounds (60.8kg). It is imperative that the installers take precautions to avoid
injury during installation.

1.

Select a mounting area with a minimum of transmitted vibration which could affect the

instrument.

2.

Verify that the mounting area is able to support the full weight of the instrument, including

options. Consider dynamic loading if the area is subject to unusual conditions.

The Purged XDS Process Analytics Instrument is shown for reference. This unit weighs 135 pounds, or
60.8 kg.