Metrohm NIRS XDS Process Analyzer – DirectLight/NonContact User Manual

Page 123

background image

▪▪▪▪▪▪▪

121

5.

Loosen the two lamp screw terminals. Slide

the lugs out from under the screws. Do not
remove the screws from the terminal strip.

6.

Twist the lamp retaining ring to the right,

and lower it from the heads of the
shoulder screws.

Remove the lamp assembly from the
instrument.

7.

Carefully remove the ring from the lamp

wires. This ring will be re-used with the
new lamp.