Warning, Digital inputs, Tou – battery replacement – SATEC PM130 PLUS Manual User Manual
Page 34

Chapter 2 Installation
I
/ O C O N N E C T I O N S
34
PM130 PLUS Powermeter Series
Digital Inputs
Figure 2-25 TOU Digital Input Connection
TOU – Battery Replacement
WARNING!
Only qualified personnel familiar with the instrument and its
associated electrical equipment must perform the RTC battery
backup replacement.
To replace the CR1632 RTC battery:
1. Remove the TOU module from the
PM130 PLUS compartment
2. Open the TOU MODULE case by applying a flat screwdriver at
three snap-in slit (1, 2 and 3), as shown in Figure 2-26.
3. Remove the old battery by lifting up the battery holder
retractable tab.
4. Place the new CR1632 battery into the battery holder while
holding up the battery holder retractable tab in such a way that
the (+) battery pole is toward the battery holder, as shown in